| Input | Type | Measurement Ranges | Min. Measurement Span | Measurement Accuracy |
| Resistance Thermometer (RTD) | Pt100 | -200°C to 850°C | 10°K | 0.1°C |
| Pt500 | -200°C to 250°C | 10°K |
| Pt1000 | -200°C to 250°C | 10°K |
| Cu50 | -50°C to 150°C | 10°K |
| Cu100 | -50°C to 150°C | 10°K |
| *Ni100 | -60°C to 180°C | 10°K |
| *Ni500 | -60°C to 180°C | 10°K |
| *Ni1000 | -60°C to 150°C | 10°K |
| Resistance Transmitter | Resistance (Ω) | 0 Ω to 400 Ω | 10 Ω | ±0.1 Ω or 0.075% |
| 0 Ω to 2000 Ω | 100 Ω | ±1.5 Ω or 0.12% |
| Thermocouples (TC) | B (PtRh30-PtRh6) | 0 to 1820°C | 500°K | Typ. 1°C |
| E (NiCr-CuNi) | -270°C to 1000°C | 50°K | 0.5°C |
| J (Fe-CuNi) | -210°C to 1200°C | 50°K |
| K (NiCr-Ni) | -270°C to 1372°C | 50°K |
| N (NiCrSi-NiSi) | -270°C to 1300°C | 50°K |
| R (PtRh13-Pt) | -50°C to 1768°C | 500°K | 1°C |
| S (PtRh10-Pt) | -50°C to 1768°C | 500°K |
| T (Cu-CuNi) | -270°C to 400°C | 50°K | 0.5°C |
| Voltage Transmitters (mV) | Millivolt Transmitter (mV) | -10 to 75mV | 5mV | 0.10% |
| -100 to 100mV | 5mV |
| -100 to 500mV | 6mV |
| -100 to 200mV | 20mV |
| OUTPUT SPECIFICATIONS |
| Output | 4 to 20mA |
| Signal on Alarm | Under Ranging | Linear Drop to 3.8mA |
| Over Ranging | Linear Rise to 20.5mA |
| Sensor break/ Open circuit | 3.6mA (Downscale) or 22mA (Upscale), selectable |
| Load | Max. (V(Power Supply) – 10.5V)/ 0.022A |
| Linearisation/ Transmission Behaviour | Temperature Linear (RTD/ T/C), Linear (Resistance or voltage inputs) |
| Galvanic Isolation | 1 KV AC (input/ output), Galvanic |
| Communiation Interface | HART Protocol (TTFUH) or PROFIBUS – PA Protocol (TTFUP) |
| PERFORMACE CHARACTERSTICS |
| Response Time | 1sec |
| Reference Operating Condition | Calibration temperature: 23°C ± 5K |
| Long Term Stability | 0.01°C (RTD), 0.1° (E,J,K,N,T), 0.2°C (B,R,S) |
| Switch on Delay | ≤ 5s |
| Influence of Ambient | Negligible |
| Filter Configuration | 0 to 160µA |
| Resolution | 0.3µA |
| ENVIRONMENT CONDITION |
| Ambent Temperature Limits | -20°C to 70°C |
| Storage Temperature | -20°C to 70°C |
| Condensation | Allowable |
| Degree of Protection | IP65 |
| Shock and Vibration Resistance | 4g/2 to 150Hz as per IEC 60 068-26 |
| Electromagnetic Compatibility (EMC) | Interference immunity and interference emission according to GB/ T17626.2-1998), compliance with IEC 61000-4-3: 1995 |
| OTHERS |
| Display Type | Visible range 32.5×22.5mm; 5 digit, 7 segment main display, digit height 8mm; 20bars meter with 5% resolution |
| Display Range | -19999 – 99999 |
| Enclosure | Dual chamber (terminals on one side & display on other) – standard single chamber (terminals below plugin display) – put suffix “S” after model number example TTFUHS |
| Weight/ Material | Approx. 800G/ Die-cast Aluminium |
| Power Supply | 9 to 32VDC, Polarity Protected |