| Supplied with 250ml Reaction Vessel, Dual Platinum Electrode, Teflon Paddle Teflon Assembly , Moisture Trap for Reaction Vessel & 250ml Amber Reservoir Bottle, Dispensing Tube (02 Nos), Teflon Tubing for Dispensing Path, Operational & Instruction Manual, Manufacturer Test Certificate and Warranty Card.Technical Specification :- Method of Detection : Conductometric/ Volumetric
- Backlit LCD Display : 20 x 4 Line Alphanumeric
- Range of Moisture Detection : less than 10 PPM to 100%
- Display of Reading : Direct Reading of Moisture in PPM, Percentage(%) & mg/ml of water
- Dispenser : Motorized Dispensing for low and accurate measurement of reagent.
- Electrode : Dual Platinum Electrode.
- Interface : Serial & Parallel Port for PC & Dot-matrix Printer attachment.
- Stirring Time : 00:00 to 99:00 ( Minutes: Second ) – Users Adjustable
- End Point Time : 1-99 Seconds , Users Adjustable
- End Point Sensitivity of Moisture Detection : Users Adjustable
- Calibration Reminder Alarm : 0-99 hours
- Data Storage : More than 100 Analysis data with corresponding calibration data.
- Input : Electrode Connector
- Power supply : 230 VAC ± 10%, 50 Hz
Salient Features :- Metallic Epoxy Based Powder Coated Body
- Parallel & Serial Port for Dot-matrix Printer & PC attachment respectively.
- Report Formation as per GLP Compliance.
- 100 analysis data storage with corresponding calibration date with Batch Number,
- Real Date & Time, Operator Name, Sample Name, searching of data as per Batch No.
- Built in Magnetic Stirrer with soft touch Key Speed Controller.
- Automatic end point detection with Audio & Visual Alarm.
- Auto Neutralization reduces the time between analysis & gives accurate Results.
- OVER TITER Audio Alarm save KF Reagent
- Automatic Calculation of Avg. Factor &10 KF Factor can be stored in memory
- Auto Drain Facility for cleaning reaction vessel.
- Onsite Dispenser Calibration / Validation Facility.
- MOC of Dispenser is SS/ABS/TEFLON for Corrosion Protection & Maintenance
- Free Operation.
- In-Built Flushing System for cleaning dispensing path.
- Users adjustable end point sensitivity.
- Calculation of % RSD
- Transfer of weight directly from Balance to KF Unit.
- RS-232 to USB interface with Cable & Driver Software to Transfer Data from Inst. to PC
Analysis Report :- Report Date & Time : 12-FEB-2010 – 16:09
- Company Name: ANALAB (VADODARA)
- Calibration Date & Time : 11-FEB-2010 – 23.34
- Calibration Method : Water by Volume
- Average Titer Factor : 6.666666
- Record Number : 00
- Analysis Date & Time : 12-FEB-2010 – 16.00
- Sample Type & Method : Liquid by Volume
- Volume of Sample (ml.)) : 05.000
- Density of Sample : 0.789
- Stirring Time (Min:Sec) : 00:00
- End Point Time (Seconds) : 20
- End Point (ml.) : 01.21
- Moisture %: 000.2044
- Moisture (PPM) : 0002044.0
- H2O (mg/ml) : 003260.33
| |
Calibration Report :- Report Date & Time : 18-JAN-2010 – 15:30
- Company Name : ANALAB (VADODARA)
- Calibration Date & Time : 18-JAN-2010 – 14.57
- Calibration Method : Water by Volume
- Stirring Time (Seconds) : 00
| |
Sr. No. | Volume(microlitre) | End Point(ml.) | TF Value | 01 | 10 | 01.83 | 5.464481 | 02 | 10 | 01.80 | 5.555556 | 03 | 10 | 01.84 | 5.434782 | 04 | 10 | 01.83 | 5.464481 | 05 | 10 | 01.85 | 5.405405 | 06 | –N/A– | | | 07 | –N/A– | | | 08 | –N/A– | | | 09 | –N/A– | | | 10 | –N/A– | | | Average Titer Factor | 5.464941 | RSD (%) | 01.02 |
|