Online Analyzer for Chlorine without reagent STAC CL
CONTEXT
It is difficult today to measure chlorine whose concentrations fluctuate from 4 mg/L up to 3 g/L. The problem still becomes complicated if measurement must be automated.
The solutions which exist to measure chlorine are:
The colorimetry using reagents and titrimetry. These methods are not easy to use on line and relate to only the small concentrations (1.5 mg/L).
Potential “Redox” measurement. Although usable in process, the measurement carried out by probe makes dilution impossible during strong concentration.
We offers a powerful and easy to use on line alternative by direct physical measurement in the UV of the chlorine concentration from 4 mg/L up to 3 000 mg/L . Much more SECOMAM Analyzer can simultaneously measure Dissolved Sulfide (H2S) in 5 – 50 mg/L range..
MEASURED PARAMETER
In this operation, H2S is dissolved at basic Ph, and then converted into Sulfate by Chlorine. The STAC Cl is able to trace both Chlorine & Sulfide.
MEASURING RANGE
It is possible to work on three ranges of measurement according to the size of the flow though cell used (2mm, 5mm)
Optical path of cell | Chlorine measuring range | Sulfide threshold detection |
2mm | From 100 to 3000 ppm | 10 ppm |
5mm | From 40 to 1200 ppm | 4 ppm |
CARACTERISTICS
Optic: UV Polychromator (204-321nm)
Samples circulation: By gravity
Maximum pressure for liquids: 1 bar
Number of measuring channels: 1 for standard analyzer (maximum 4)
Measurement frequency: 4 minutes per channel
Memory: 4 models, 255 results
Working temperature: From 15 to 40°C
Samples temperature: 40°C maximum
Nature of the samples Natural or treated (tparticules size< 2mm)
Outputs: 4 current loops (4-20mA) and RS 232C port
Consumable: Demineralized or distilled water (blank)
Source: Deuterium lamp (100 000 measures maximum)
Power: Supply 230V single phase – 50/60Hz – 300VA
Size :800 x 600 x 360
Weight: 15 Kg