Karl Fischer Titrator Model : OptiCal100

 

 

Supplied with 250ml Reaction Vessel, Dual Platinum Electrode, Teflon Paddle Teflon Assembly , Moisture Trap for Reaction Vessel & 250ml Amber Reservoir Bottle, Dispensing Tube (02 Nos), Teflon Tubing for Dispensing Path, Operational & Instruction Manual, Manufacturer Test Certificate and Warranty Card.

Technical Specification :

  • Method of Detection : Conductometric/ Volumetric
  • Backlit LCD Display : 20 x 4 Line Alphanumeric
  • Range of Moisture Detection : less than 10 PPM to 100%
  • Display of Reading : Direct Reading of Moisture in PPM, Percentage(%) & mg/ml of water
  • Dispenser : Motorized Dispensing for low and accurate measurement of reagent.
  • Electrode : Dual Platinum Electrode.
  • Interface : Serial & Parallel Port for PC & Dot-matrix Printer attachment.
  • Stirring Time : 00:00 to 99:00 ( Minutes: Second ) – Users Adjustable
  • End Point Time : 1-99 Seconds , Users Adjustable
  • End Point Sensitivity of Moisture Detection : Users Adjustable
  • Calibration Reminder Alarm : 0-99 hours
  • Data Storage : More than 100 Analysis data with corresponding calibration data.
  • Input : Electrode Connector
  • Power supply : 230 VAC ± 10%, 50 Hz

Salient Features :

  • Metallic Epoxy Based Powder Coated Body
  • Parallel & Serial Port for Dot-matrix Printer & PC attachment respectively.
  • Report Formation as per GLP Compliance.
  • 100 analysis data storage with corresponding calibration date with Batch Number,
  • Real Date & Time, Operator Name, Sample Name, searching of data as per Batch No.
  • Built in Magnetic Stirrer with soft touch Key Speed Controller.
  • Automatic end point detection with Audio & Visual Alarm.

 

  • Auto Neutralization reduces the time between analysis & gives accurate Results.
  • OVER TITER Audio Alarm save KF Reagent
  • Automatic Calculation of Avg. Factor &10 KF Factor can be stored in memory
  • Auto Drain Facility for cleaning reaction vessel.
  • Onsite Dispenser Calibration / Validation Facility.
  • MOC of Dispenser is SS/ABS/TEFLON for Corrosion Protection & Maintenance
  • Free Operation.
  • In-Built Flushing System for cleaning dispensing path.
  • Users adjustable end point sensitivity.
  • Calculation of % RSD
  • Transfer of weight directly from Balance to KF Unit.
  • RS-232 to USB interface with Cable & Driver Software to Transfer Data from Inst. to PC

Analysis Report :

  • Report Date & Time : 12-FEB-2010 – 16:09
  • Company Name: ANALAB (VADODARA)
  • Calibration Date & Time : 11-FEB-2010 – 23.34
  • Calibration Method : Water by Volume
  • Average Titer Factor : 6.666666
  • Record Number : 00
  • Analysis Date & Time : 12-FEB-2010 – 16.00
  • Sample Type & Method : Liquid by Volume
  • Volume of Sample (ml.)) : 05.000
  • Density of Sample : 0.789
  • Stirring Time (Min:Sec) : 00:00
  • End Point Time (Seconds) : 20
  • End Point (ml.) : 01.21
  • Moisture %: 000.2044
  • Moisture (PPM) : 0002044.0
  • H2O (mg/ml) : 003260.33

 

Calibration Report :

  • Report Date & Time : 18-JAN-2010 – 15:30
  • Company Name : ANALAB (VADODARA)
  • Calibration Date & Time : 18-JAN-2010 – 14.57
  • Calibration Method : Water by Volume
  • Stirring Time (Seconds) : 00

 

Sr. No. Volume(microlitre) End Point(ml.) TF Value
01 10 01.83 5.464481
02 10 01.80 5.555556
03 10 01.84 5.434782
04 10 01.83 5.464481
05 10 01.85 5.405405
06 –N/A–    
07 –N/A–    
08 –N/A–    
09 –N/A–    
10 –N/A–    
Average Titer Factor 5.464941
RSD (%) 01.02